| Details
Current measurement range from 300 fA to 10
µA (four ranges) permits wide dynamic range for detection * Wide
bandwidth allows recording of fast events * Measure nitric oxide from <
0.3 nM to 100 µM * Measure hydrogen peroxide < 10 nM to 100 mM *
Measure hydrogen sulfide * Measure glucose * Measure oxygen from 0.1% to
100% * Isolated architecture allows Lab-Trax interface to simultaneously
measure free radical and independent analog data (i.e., ECG, BP, etc.)
data on any channel
Real-time detection
Real-time detection and measurement of a variety
of redox-reactive species is fast and easy using the electrochemical (amperometric)
detection principle employed in the new TBR4100. This optically isolated
four-channel free radical analyzer has ultra low noise and independently
operated channels.
Measure multiple species
For use with WPI’s wide range of nitric oxide,
hydrogen peroxide, hydrogen sulfide and oxygen sensors, the TBR4100 can
measure four different species simultaneously in the same preparation.
Simply plug a sensor into any one of the input channels on the front panel
and select the current range. Poise voltage can be selected from a range
of values tuned for optimal response from WPI sensors. An independent
output for real-time monitoring of temperature is also included.
Data acquisition
The TBR4100 analyzer utilizes PC-based data
acquisition via our Lab-Trax interface; data traces are displayed and
recorded in real-time. The LabScribe software comes pre-configured for
single or multiple electrode recording; filters, gains, and smoothing are
all set for optimal results. Data can be viewed making adjustments to
smoothing and filter settings without affecting the original stored raw
data. Electrode calibration from multiple concentration readings can be
input into the software's Multipoint Calibration utility quickly provides
a plot and slope calculation for electrode sensitivity determination.
Alternately, the Lab-Trax T series data interface can be used for
providing simultaneous acquisition of Free Radical data along with other
physiological data (ECG, HR, BP, etc.) as each of the four input channels
has its own independent input amplifier, filters, and 24-bit converter.
Turnkey systems
TBR4100-416 includes TBR4100 analyzer & power
cord, Lab-Trax-4/16 data acquisition system & USB cable, 4 BNC cables,
3 electrode adapter cables, 1 temperature probe, 2 sensors of your choice,
and sensor start-up kit(s), if applicable * TBR4100-424T includes TBR4100
analyzer & power cord, Lab-Trax-4/24T data acquisition system &
USB cable, 4 BNC cables, 3 electrode adapter cables, 1 temperature probe,
2 sensors of your choice, and sensor start-up kit(s), if applicable.
For more information on Lab-Trax, including
screen shots, see the article.
For more information on features available in the Data-Trax software for
Lab-Trax (now called LabScribe), see the article.
| Power |
100 ~ 240 VAC, 50-60 Hz, <15 W |
| Operating Temperature (ambient) |
0 - 50°C (32 - 122°F) |
| Operating Humidity (ambient) |
15 - 70% RH non-condensing |
| Warm up Time |
<5 minutes |
| Dimensions |
135 X 419 X 217 mm (5.25" X
16.5" X 8.16") |
| Weight |
1.35 kg (3 lb) |
| Display Functions |
18 mm (0.7") LCD readout, 4.5 digit
Polarization Voltage (mV) Current input (nA, µA) |
| Controls |
Power (on/off) Current Input Range
Polarization Voltage |
| Analog Output Range |
+/- 10 V (continuous) |
| Analog Output Impedance |
10 kohm |
| Channel to Channel Isolation |
>10 Gohm |
| Channel to Output Isolation |
>10 Gohm |
| Power Supply to AC Line Isolation |
>100 Mohm |
| Analog Output Drift |
<10 pA/h |
| Temperature Input: Number of Channels |
1 |
| Temperature Input: Sensing Element |
Platinum RTD, 1000 Ohm |
| Temperature Input: Range |
0-100°C |
| Temperature Input: Accuracy |
+/- 1°C |
| Temperature Input: Resolution |
0.1°C |
| Temperature Input: Analog Output |
31.25 mV/°C (continuous) |
| Amperometric Input: Number of Amperometric
Channels |
4 |
| Amperometric Input: Signal Bandwidth |
0-3 Hz |
| Amperometric Input: Polarization Voltage
(selectable via rotary switch) Nitric Oxide |
865 mV |
| Amperometric Input: Polarization Voltage
(selectable via rotary switch) Hydrogen Sulfide |
150 mV |
| Amperometric Input: Polarization Voltage
(selectable via rotary switch) Hydrogen Peroxide |
450 mV |
| Amperometric Input: Polarization Voltage
(selectable via rotary switch) Glucose |
600 mV |
| Amperometric Input: Polarization Voltage
(selectable via rotary switch) Oxygen |
700 mV |
| Amperometric Input: Polarization Voltage
(selectable via rotary switch) ADJ (user adjustable) |
+/- 2500 mV |
| Polarization Voltage Accuracy |
+/- 5 mV |
| Polarization Voltage Display Resolution |
+/- 1mV |
| Current measurement Performance: Range |
+/- 10 Na, +/- 100 nA, +/- 1
µA, +/- 10 µA |
| Current measurement Performance: Analog
Output |
1 mV / 1 pA, 1 mV / 10pA, 1 mV / 100pA, 1
mV / 1µA |
| Current measurement Performance: Noise @
3Hz * |
< 1 pA, < 7 pA, < 70 pA, < 700
pA |
| Current measurement Performance: Noise @
0.3 Hz * |
< 0.3 pA, < 3 pA, < 30 pA, <
300 pA |
| Notes: |
*Instrument performance is measured as the
(max-min) over 20 seconds period with open input. Typical values
are given at 3 Hz and 0.3 Hz bandwidth. |
| Typical sensor performance with TBR4100:
ISO-NOPF100 noise |
0.2 nM NO (<2 pA)** |
| Notes: |
**Sensor noise is measured as the
(max-min) over a 20 seconds period with the sensor immersed in 0.1
M CuCl2 solution. |
91580 - microsensor adapter cable
ISO-TEMP-2 - Temperature Sensor
ISO-HPO-100 - Hydrogen Peroxide Sensor
ISO-HPO-2 - Hydrogen Peroxide Sensor
ISO-H2S-2 - Hydrogen Sulfide Sensor
ISO-GLU - Implatable Glucose Sensor
ISO-NOP - Nitric Oxide Sensor
ISO-HPO-100-L - Hydrogen Peroxide Sensor - 100um
ISO-HPO-100-H - Hydrogen Peroxide Sensor - 100um with hypodermic sheath
ISO-NOPNM - Nitric Oxide Sensor - 100nm
ISO-NOPF100 - Nitric Oxide Sensor - 100um
ISO-NOPF200-L10 - Nitric Oxide Sensor - 200um
ISO-NOPF200 - Nitric Oxide Sensor - 200um
ISO-NOP3020 - Nitric Oxide Sensor - 30 micron
ISO-NOP3005 - Nitric Oxide Sensor - 30 micron
ISO-NOP30-L - Nitric Oxide Sensor - 30 micron
ISO-NOP007 - Nitric Oxide Sensor - 7 micron
ISO-NOP-70-L - Nitric Oxide Sensor - 70 micron
ISO-OXY-2 - Oxygen Sensor - 2mm
|